ACL-2 Colorimeter For Petroleum Products - TANAKA[Mar. 10, 2010 3:02:37]
Specification
The ACL Colorimeter was designed for automatic Saybolt/ ASTM Color measurement of petroleum products utilizing the Tri-Stimulus method. No subjective judgment, plus full capability of easy and precise calibration, assures a high level traceability required by ISO 9000 series programs. Combination of high speed 16 bit A/ D converter and a low drift receiver assures extremely high accuracy. Test standards ASTM D 6045, JIS K2580.
* Tri-Stimulus Method
* High Accuracy
Traceability: No subjective judgment plus full capability of easy and precise calibration assure the high level of traceability that is required by modern quality assurance programs such as the ISO-9000 series.
Accuracy: The combination of a high-speed 16 bit A/ D converter and low drift/ high precision amplifier assures extremely high accuracy. No more guess work, especially for the tricky Saybolt color measurement.
Quicker than Spectrophotometer: A typical measurement takes a mere 9 seconds, which is considerably shorter than the time a spectrophotometer would require. ( typically 25 seconds)
User Friendly Interface:
* A dialogue type operation plus dedicated design for color measurement of petroleum products assure the easy operation. Just select the desired color scale from the menu, pour a sample in the sample cell, set it in the cell compartment, and press the START key; an accurate test result will be ready in 7-10 seconds.
* The new design sample cell takes a smaller sample volume ( 1/ 3 of the conventional rectangular cell) because of its cylindrical design. Also featured are plastic caps, which prevent sample spill when the sample cell is set in the instrument.
* Capabilities of Platinum-Cobalt color scale for solvents and Gardner color scale for varnish have been added.
Wide Selection of Options:
* Flow-through cells for easy connection to automated sample handling systems.
* Calibration Fluid Samples made according to the Appendix of ASTM D6045 for easy/ accurate calibration.
* Rectangular sample cells for easy pouring of wax samples.
Specifications
Type:
Transmittance colorimeter for petroleum products by tri-stimulus method.
Conforming Standards:
ASTM D6045, JIS K2580
Color Series:
Saybolt color ( ASTM D6045/ D156) ,
ASTM color ( ISO 6271, ASTM D6045/ D1500) ,
Platinum-Cobalt color ( ISO 6271, ASTM D1209)
Gardner color ( ISO 4630, ASTM D1544)
Light Source:
Halogen bulb ( 12V 20W)
Light Receiver:
Fast response silicon photo cells
Tri-Stimulus Filters:
Colored glass filters
Displayed Parameters:
Saybolt color, L* a* b, DL* , Da* , Db* , DE*
ASTM color, X, Y, Z, DX, DY, DZ, £ D
Pt-Co color YI Gardner color
Automatic Calibration:
Single point or multi point calibration using calibration samples.
Sample Cells:
Cylindrical glass cell with caps
100mm pass length for Saybolt color
33mm pass length for ASTM, Pt-Co, Gardner color
Display Module:
Fluorescent display module, 256 x 64 dots
Printer:
Thermal printer with 80mm paper width
Computer Interface:
1 set of RS-232C port ( standard)
Dimensions & Weight:
400W( mm) x 400D( mm) x 190H( mm) , 12.2Kg
Power Consumption:
AC 100-240V, 130VA