Specification :
DESCRIPTION:
Fully automatic,random access, continuous loading, benchtop analyzer for Clinical Chemistry and Immunoturbidimetric assays.
ASSAY TIPE:
End point, Kinetic, Differential, Bichromatic and Immunoturbidimetric; up to 999 tests, all user programmable.
ON LINE REAGENTS:
4 racks are provided to hold up to 33 reagents that are divided as follows:
- 2 racks cooled by Peltier cells holding up to 20 reagents
- 2 racks for 13 reagent positions and 14 standards and controls at room temperature
- 2 additional positions for diluents.
SAMPLES LOADING:
5 racks for continuous loading of samples split in:
- 64 positions for routine samples
- 14 positions for STAT samples
PRIMARY TUBES SIZE:
max 16 mm x 100 mm (height)
STAT SAMPLES PROCESSING:
Allowed at any time
THROUGHPUT:
Up to 333 test/hour with ISE (200 without ISE)
WORKING TEMPERATURE:
37°C
READING SYSTEM:
Direct photometry
MINIMUM REACTION VOLUME:
300 µl
OPTICS SYSTEM:
- Photometer: multi - wavelenght dual channel optic
- Wavelenght: 8 narrow band interferential filters
- Light Source: halogen lamp 6V/10W
- Optical Path: 10 mm.
- Linear Range: 0.001 ÷ 2.500 Abs
- Resolution: 0.0001 Abs
SAMPLING ARM:
One mechanical arm performs all sampling operation with:
- capacitive liquid level sensing
- reagent pre-warming at 37°C
- automatic probe washing
DILUTER:
Dilutions are made without using traditional syringes with the following specifications:
- sample volume: 3µl ÷ 99 µl (1 µl incr.)
- reagent volume: 3µl ÷ 500 µl (1 µl incr.)
REAGENT PLATE:
Circular 60 semi-disposable cuvettes tray with the following carachteristics:
- automatic cuvette washing
- cuvettes Q.C. continuously monitored by the computer
- incubation temperature: 37°C
COMPUTER (miminum requirements):
- CPU: Pentium III, 500 Mhz
- MEMORY: 256 Mb
- CD Rom: 32X
- FD: 1.44 Mb
- HD: 20 Gb IDE
- Printer: Epson LX 300+
- Communication: 2 RS232 ports, 1 parallel port
- Software: WINDOWS 98
OPTIONS
- Direct potentiometry ISE module for Na+, K+, Cl
- Positive barcode reader
INSTALLATION REQUIREMENTS
Power Supply:
- 90 ÷ 250 Vac
- 47/63 Hz
- 300 W
Water Consumption:
2 ml/test
DIMENSION, WEIGHT & ENVIRONMENT
Analytical Module:
- Height: 42 cm
- Depht: 65 cm
- Lenght: 100 cm
- Weight 65 Kg
Room Temperature:
18° C ÷ 30° C
Relative Humidity:
20% ÷ 85%