THERMOCOUPLE THERMOMETERS: K, J, T, N, R, S, B, E, Merk : DeltaOhm
Specification
Thermocouple Thermometers
HD2108.1 and HD2108.2 with one input HD2128.1 and HD2128.2 with two inputsThe HD2108.1 and HD2108.2 with one input and the HD2128.1 and HD2128.2 with two inputs are portable instruments with a large LCD display.
They measure the temperature using immersion, penetration air or contact probes. The sensor may be a thermocouple of type K, J, T, N, R, S, B or E. The HD2108.2 and HD2128.2 instruments are dataloggers. The HD2108.1 memorizes up to 76,000 samples, the HD2128.2 up to 38,000 pairs of values. These data can be transferred from the instrument connected to a PC via the multi-standard RS232C serial port and USB 2.0. The storing interval, printing, and baud rate can be confi gured using the menu.All the models are fi tted with an RS232C serial port and can transfer the acquired measurements in real time to a PC or to a portable printer.The Max, Min and Avg function calculate the maximum, minimum or average values. Other functions include: the relative measurement REL, the HOLD function, and the automatic turning off that can also be disabled. The HD2128.1 and HD2128.2 calculate the A-B difference of the temperatures captured by the two input channels. The instruments have IP67 protection degree.
INSTRUMENT TECHNICAL CHARACTERISTICS
Instrument
Dimensions (Length x Width x Height) 185x90x40mm
Weight 470g (complete with batteries)
Materials ABS, rubber
Display 2x41D2 digits plus symbols
Visible area: 52x42mm
Operating conditions
Operating temperature -5…50°C
Warehouse temperature -25…65°C
Working relative humidity 0…90%RH without condensation
Protection degree IP67
Power Batteries 4 1.5V type AA batteries
Autonomy 200 hours with 1800mAh alkaline batteries
Power absorbed with instrument off 20¼A
Mains Output mains adapter 9Vdc / 250mA
Measuring unit °C - °F - °K - mV - mV*C
Security of memorized data Unlimited, independent of battery charge conditions
Time Date and time
Schedule in real time
Accuracy 1min/month max departure
Measured values storage
Type - model HD2108.2 2000 pages containing 38 samples each
Total of 76,000 samples
Type - model HD2128.2 2000 pages containing 19 samples each 38,000 pairs of samples
Storage interval 1s…3600s (1hour)
Serial interface RS232C
Type RS232C electrically isolated
Baud rate Can be set from 1200 to 38400 baud
Data bit 8 Parity None Stop bit 1
Flow Control Xon/Xoff
Serial cable length Max 15m
Immediate print interval 1s…3600s (1hour)
USB interface - model HD2108.2 and HD2128.2
Type 1.1 - 2.0 electrically isolated
Connections Input for probes 2-pole female polarized standard miniature connector
Serial interface and USB 8-pole MiniDin connector
Mains adapter 2-pole connector (positive at centre)
Measurement of temperature by Instrument TC
measurement range: K -200…+1370°C TC
measurement range: J -100…+750°C TC
measurement range: T -200…+400°C TC
measurement range: N -200…+1300°C TC
measurement range: R +200…+1480°C TC
measurement range: S +200…+1480°C TC
measurement range: B +200…+1800°C TC
measurement range: E -200…+750°C
Resolution 0.05°C up to 199.95°C 0.1°C from 200.0°C to the full scale
Instrument precision
Thermocouple K ±0.1°C up to 600°C ±0.2°C over 600°C Thermocouple J ±0.05°C up to 400°C ±0.1°C over 400°C Thermocouple T ±0.1°C
Thermocouple N ±0.1°C up to 600°C ±0.2°C over 600°C Thermocouple R ±0.25°C
Thermocouple S ±0.3°C
Thermocouple B ±0.35°C
Thermocouple E ±0.1°C up to 300°C ±0.15°C over 300°C
The accuracy only refers to the instrument. Error due to the thermocouple or to the cold junction reference sensor is not included.
Temperature drift @ 20°C 0.02%/°C
Drift after 1 year 0.1°C/year
Accuracy of the thermocouple probes :
The tolerance of a type of thermocouple corresponds to the maximum acceptable shift from the e.m.f. of any thermocouple of that type, with reference junction at 0°C. The tolerance is expressed in degrees Celsius, preceded by the sign. The percentage tolerance is given by the ratio between the tolerance expressed in degrees Celsius and the measurement junction temperature, multiplied by one hundred. The thermocouples conforming to regulations must comply with one of the following tolerance levels, the values of which are reported in the table.
G I (special tolerances)
G II (normal tolerances)
The tolerances refer to the operating temperature expected for the thermocouple, in agreement with the thermoelements’ diameter.
ORDER CODES
HD2108.1K: The kit is composed of the instrument HD2108.1 with one input, connection cable for serial output HD2110CSNM, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. The probes must be ordered separately.
HD2108.2K: The kit is composed of the HD2108.2 with one input, datalogger, connection cable HD2101/USB, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. The probes must be ordered separately.
HD2128.1K: The kit is composed of the instrument HD2128.1 with two inputs, connection cable for serial output HD2110CSNM, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. The probes must be ordered separately.
HD2128.2K: The kit is composed of the HD2128.2 with two inputs, datalogger, connection cable HD2101/USB, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. The probes must be ordered separately.
HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole female for RS232C.
HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole MiniDin.
DeltaLog9: Software for download and management of the data on PC using Windows 98 to XP operating systems.
AF209.60: Stabilized power supply at 230Vac/9Vdc-300mA mains voltage.
S’print-BT: On request, portable, serial input, 24 column thermal printer, 58mm paper width.
Thermocouple probes
The instruments can be connected to all the thermocouple probes fi tted with standard miniature connector available on our price-list.