    Waveform Generators Cat.No43944

    Quantity Order:
    Payment Method:
    Telegraphic Transfer (T/T)

    Specification :

    The 290 series uses Direct Digital Synthesis techniques as well as variable clock sampling technology to provide a fully featured programmable function and arbitrary waveform capability.

    The 291, 292 and 294 are 100 MS/ s arbitrary waveform generators designed to handle real world requirements by test experts. These models can easily simulate complex signals while being easy to use, compact and affordable. On multi-channel units each channel can be operated fully independently, or multiple channels can be linked using simple or complex relationships.

    Waveform Manager Plus software provides all the features needed for creation, manipulation and management of arbitrary waveforms within a single Windows based program.
    Arbitrary Waveforms
    Waveforms The maximum arbitrary waveform size is 1 M points. Up to 500 user-defined waveforms may be stored on the removable memory card. Arbitrary waveforms can be defined by front panel editing controls, by downloading of waveform data via RS-232, USB or GPIB, or by writing directly to the removable memory card using the USB card reader/ writer connected to a PC.

    Waveform memory 1 M points. Minimum waveform size is 8 points.

    Vertical resolution 12 bits

    Sample clock range 100 mHz to 100 MHz

    Resolution 4 digits

    Accuracy ± 1 digit of setting

    Output filter Selectable between 40 MHz Elliptic, 20 MHz Bessel or none

    Sequence Up to 1024 waveforms may be linked. Each waveform can have a loop count of up to 32, 768. A sequence of waveforms can be looped up to 1, 048, 575 times or run continuously.

    Noise function Digital noise generated by a 35-bit linear feedback register clocked at 100 MHz. User’ s external filter defines bandwidth and response

    Standard Waveforms
    Waveforms Sine, square, triangle, DC, positive ramp, negative ramp, sin( x) / x, pulse, pulse train, cosine, haversine and havercosine

    Sine, Cosine, Haversine, Havercosine
    Range 0.1 mHz to 40 MHz

    Resolution 0.1 mHz or 10 digits

    Accuracy Better than 10 ppm for 1 year

    Temperature stability Typically < 1 ppm/ ° C

    Output level 5 mV to 20 V p-p from 50 &

    Harmonic distortion < 0.15 % THD to 100 kHz; < – 60 dBc to 20 kHz, < – 50 dBc to 1 MHz, < – 40 dBc to 10 MHz, < – 30 dBc to 40 MHz

    Non-harmonic spurii < – 60 dBc to 1 MHz, < – 60 dBc + 6 dB/ octave 1 MHz to 40 MHz

    Square
    Range 1 mHz to 50 MHz

    Resolution 1 mHz ( 4 digits)

    Accuracy ± 1 digit of setting

    Output level 5 mV to 20 V p-p from 50 &

    Rise and fall times < 8 ns

    Triangle
    Range 0.1 mHz to 500 kHz

    Resolution 0.1 mHz or 10 digits

    Accuracy Better than 10 ppm for 1 year

    Output level 5 mV to 20 V p-p from 50 & , linearity error: < 0.1 % to 30 kHz

    Ramps and Sin( x) / x
    Range 0.1 mHz to 500 kHz

    Resolution 0.1 mHz or 10 digits

    Accuracy Better than 10 ppm for 1 year

    Output level 5 mV to 20 V p-p from 50 &

    Linearity error < 0.1 % to 30 kHz

    Pulse and Pulse Train
    Output level 5 mV to 20 V p-p from 50 &

    Rise and fall times < 8 ns

    Period Range: 40 ns to 100 s; Resolution: 4-digits; Accuracy: ± 1 digit of setting

    Delay Range: – 99.9 s to + 99.99 s; Resolution: 0.001 % of period or 10 ns

    Width Range: 10 ns to 99.99 s; Resolution: 0.001 % of period or 10 ns

    Trains of up to 10 pulses may be specified, each having independently defined width, delay and level. The baseline voltage is separately defined and the sequence repetition rate is set by the pulse train period.
    Operating Modes
    Continuous Waveform runs continuously

    Triggered Burst Each active edge of the trigger signal will produce one burst of the waveform

    Carrier waveforms All standard and arbitrary

    Max. carrier frequency The smaller of 2.5 MHz or the maximum for the selected waveform. 100 Msamples/ s for ARB or Sequence.

    Number of Cycles 1 to 1048575

    Trigger rep. rate 0.005 Hz to 100 kHz internal, dc to 1 MHz external

    Trigger source Internal from keyboard or trigger generator. External from TRIG IN or remote interface.

    Start/ stop phase ± 360 ° settable with 0.1 ° resolution, subject to waveform frequency and type

    Gated Waveform will run while the Gate signal is true and stop while false

    Carrier waveforms All standard and arbitrary

    Max. carrier frequency The smaller of 2.5 MHz or the maximum for the selected waveform. 80 Msamples/ s for ARB or Sequence.

    Trigger rate 0.005 Hz to 100 kHz internal, dc to 1 MHz external

    Gate signal source Internal from keyboard or trigger generator. External from TRIG IN or remote interface.

    Start/ stop phase ± 360 ° settable with 0.1 ° resolution, subject to waveform frequency and type

    Sweep Capability provided for both standard and arbitrary waveforms. Arbitrary waveforms are expanded or condensed to exactly 4096 points and DDS techniques are used to perform the sweep.

    Carrier waveforms All standard and arbitrary except pulse, pulse train and sequence. Sweep mode: Linear or logarithmic, continuous or triggered.

    Sweep direction Up, down, up/ down or down/ up

    Sweep range 1 mHz to 40 MHz in one range. Phase continuous. Independent setting of start/ stop frequency.
    http: / / www.midwest-g.com
    Sweep time 1 ms to 999 s ( 3 digit resolution)

    Marker Variable during sweep.

    Sweep trig. Source The sweep may be free run or triggered from the following sources: Manually from keyboard. Externally from TRIG IN input or remote interface.

    Sweep hold Sweep can be held and restarted by HOLD key

    Tone Switching Capability provided for both standard and arbitrary waveforms. Arbitrary waveforms are expanded or condensed to exactly 4096 points and DDS techniques used to allow instantaneous frequency switching.

    Carrier waveforms All waveforms bar pulse, pulse train, sequence

    Frequency list Up to 16 frequencies from 1 mHz to 40 MHz

    Trigger rep. rate 0.005 Hz to 100 kHz internal, dc to 1 MHz external. Usable repetition rate and waveform frequency depend on the tone switching mode.

    Source Internal from keyboard or trigger generator. External from TRIG IN or remote interface.

    Tone switching modes

    Gated The tone is output while the trigger signal is true and stopped, at the end of the current waveform cycle, while the trigger signal is false. The next tone is output when the trigger signal is true again.

    Triggered The tone is output when the trigger signal goes true and the next tone is output, at the end of the current waveform cycle, when the trigger signal goes true again.

    FSK The tone is output when the trigger signal goes true and the next tone is output, immediately, when the trigger signal goes true again.

    External Amplitude Modulation
    Carrier frequency Entire range for selected waveform

    Carrier waveforms All standard and arbitrary waveforms

    Modulation source Modulation socket

    Frequency range DC to 500 kHz

    Signal range Approx. 1 V pk-pk for 100 % level change at maximum output

    External Signal Summing
    Carrier frequency Entire range for selected waveform

    Carrier waveforms All standard and arbitrary waveforms

    Sum source Sum socket

    Frequency range DC to 16 MHz

    Signal range Approximately 2 Vpk-pk input for 20 Vpk-pk output.

    Trigger Generator
    Source Internal source 0.005 Hz to 100 kHz squarewave adjustable in 10 us steps. 3 digit resolution. Available for external use from the SYNC OUT socket.

    Main Outputs – One for each channel
    Output impedance 50 &

    Amplitude 5 mV to 20 V pk-pk open circuit ( 2.5 mV to 10 V pk-pk into 50 & . Amplitude can be specified open circuit ( Hi Z) or into an assumed load of 50 & or 60 & , in Vpk-pk, Vrms or dBm.

    Ampl. accuracy Better than 2 % ± 1 mV at 1 kHz into 50 & .

    Ampl. flatness ± 0.2 dB to 1 MHz; ± 0.4 dB to 40 MHz

    DC offset range ± 10 V. DC offset plus signal peak limited to ± 10 V from 50 & .

    Offset accuracy Typically within ± 3 % ± 10 mV, unattenuated

    Resolution 3 digits or 1 mV for both Amplitude and DC Offset

    Sync Out – One for each channel Multifunction output user definable or automatically selected to be any of the following:

    Waveform sync ( All waveforms) A square wave with 50 % duty cycle at the main waveform frequency, or a pulse coincident with the first few points of an arbitrary waveform.

    Position markers ( Arbitrary only) Any point( s) on the waveform may have associated marker bit( s) set high or low

    Burst done Produces a pulse coincident with the last cycle of a burst

    Sequence sync Produces a pulse coincident with the end of a waveform sequence

    Trigger Selects the current trigger signal. Useful for synchronising burst or gated signals.

    Sweep sync Outputs a pulse at the start of sweep to synchronise an oscilloscope or recorder. Can additionally output a sweep marker.

    Phase lock out Used to phase lock two generators. Produces a positive edge at the 0o phase point.

    Output signal level Logic level of 3 V for all outputs except Sweep Sync. Sweep Sync is a 3-level waveform.

    Trig In
    Frequency range DC to 1 MHz

    Signal range Threshold nominally TTL level; max. input ± 10 V

    Min. rulse width 50 ns for Trigger and Gate modes; 50 µ s for Sweep mode

    Input impedance 10 k&

    Modulation In
    Frequency range DC to 500 kHz

    Signal range VCA: Approximately 1 Vpk-pk for 100 % level change at maximum output
    SCM: Approximately ± 1 Vpk for maximum output

    Input impedance Typically 1 k&

    Sum In
    Frequency range DC to 30 MHz ( 291) DC to 16 MHz ( 292/ 294)

    Signal range Approximately 2 Vpk-pk input for 20 Vpk-pk output

    Input impedance Typically 1 k&

    Hold Holds an arbitrary waveform at its current position. A TTL low level or switch closure causes the waveform to stop at the current position and wait until a TTL high level or switch opening which allows the waveform to continue. The front panel MAN/ HOLD key or remote command may also be used to control the Hold function.

    Input impedance 10 k&

    Ref Clock In/ Out
    Set to input Input for an external 10 MHz reference clock. TTL/ CMOS threshold level.

    Set to output Buffered version of the internal 10 MHz clock. Output levels nominally 1 V and 4 V from 50 & .

    Set to phase lock Used together with SYNC OUT on a master and the TRIG IN on a slave to synchronize ( phase lock) two generators

    ARB Clock In
    Frequency Range DC to 50 MHz

    Max. input voltage + 5 V, -1 V

    Inter-Channel Operation
    Inter-Channel Modulation The waveform from any channel may be used to Amplitude Modulate ( AM) or Suppressed Carrier Modulate ( SCM) the next channel. Alternatively any number of channels may be Modulated ( AM or SCM) with the signal at the MODULATION input socket.

    Carrier frequency Entire range for selected waveform

    Carrier waveforms All standard and arbitrary waveforms

    Modulation types AM: Double sideband with carrier
    SCM: Double sideband suppressed carrier

    Modulation source Internal from the previous channel. External from Modulation input socket. The external modulation signal may be applied to any number of channels simultaneously

    Frequency range DC to > 100 kHz

    Internal AM depth 0 % to 105 %

    Internal AM resolution 1 %

    Carrier Suppression ( SCM) > 40 dB

    External modulation signal range VCA: Approximately 1 V pk-pk for 100 % level change at maximum output
    SCM: Approximately ± 1 Vpk for max. output

    Inter-Channel Analogue Summing Waveform Summing sums the waveform from any channel into the next channel. Alternatively any number of channels may be summed with the signal at the SUM input socket.

    Carrier frequency Entire range for selected waveform

    Carrier waveforms All standard and arbitrary waveforms

    Sum source Internal from the previous channel. External from SUM IN socket.

    Frequency range DC to > 16 MHz

    Ext. signal range Approx. 5 Vpk-pk input for 20 Vpk-pk output

    Inter-Channel Phase Locking Two or more channels may be phase locked together. Each locked channel may be assigned a phase angle relative to the other locked channels. Arbitrary waveforms and waveform sequences may be phase locked but certain constraints apply to waveform lengths and clock frequency ratios.
    With one channel assigned as the Master and other channels as Slaves a frequency change on the master will be repeated on each slave thus allowing multiphase waveforms at the same frequency to be easily generated. DDS waveforms are those with 7 digits of frequency setting resolution, while Non-DDS waveforms have 4 digits.

    Phase resolution DDS waveforms: 0.1 degree

    Non-DDS waveforms 0.1 degree or 360 degrees/ number of points whichever is the greater

    Phase error < ± 10 ns all waveforms.

    Inter-Channel Triggering Any channel can be triggered by the previous or next channel. The previous/ next connections can be used to ‘ daisy chain’ a trigger signal from a ‘ start’ channel, through a number of channels in the ‘ chain’ to an ‘ end’ channel.
    Each channel receives the trigger out signal from the previous ( or next) channel, and drives its selected trigger out to the next ( or previous) channel. The ‘ end’ channel trigger out can be set up to drive the ‘ start’ channel, closing the loop.
    In this way, complex and versatile interchannel trigger schemes may be set up. Each channel can have its trigger out and its output waveform set up independently. Trigger out may be selected from Waveform End, Position Markers, Sequence Sync or Burst Done.
    Using the scheme above it is possible to create a sequence of up to 64 waveform segments, each channel producing up to 16 segments and all channels being summed to produce the complete waveform at the output of channel 4.
    The signals from the REF IN/ OUT socket and the SYNC OUT socket can be used to phase lock two instruments where more than 4 channels are required.

    Interfaces
    RS-232 Variable Baud rate, 9600 Baud maximum

    IEEE488 Conforms with IEEE488.2

    USB Conforms with USB 1.1

    General Specifications
    Display 20 character x 4 row alphanumeric LCD

    Data entry Keyboard selection of mode, waveform etc.; value entry by numeric keys or by rotary control.

    Memory card Removable memory card conforming to the Compact Flash memory card standard. Sizes from 32 MB to 1 GB can be used.

    Stored settings Up to 500 complete instrument set-ups may be stored and recalled from the memory card. Up to 500 arbitrary waveforms can also be stored independent of the instrument settings.

    Size 130 mm ( 3U) high; 335 mm long; 350 mm wide ( 292/ 294) , 212 mm wide ( 291)

    Weight 292/ 294: 7.2 kg ( 16 lb) ; 291: 4.1 kg ( 9 lb)

    Power 110 to 120 V or 100 V nominal 50/ 60/ 400 Hz; 220 to 240 V nominal, 50/ 60 Hz. Voltage adjustable internally; operating range ± 10 % of nominal; 60 VA max. Installation Category II.

    Compliance
    Operating range + 5 ° C to 40 ° C, 20 to 80 % RH

    Storage range – 20 ° C to + 60 ° C

    Environmental Indoor use at altitudes to 200 m, Pollution Degree 2

    Safety Complies with EN61010-1

    EMC Complies with EN61326

    Instrument drivers Labview and LabWindows CVI drivers are either supplied with the instrument or are available via your local Fluke Office

    Supplied Items IEC Mains Lead. Printed manual ( partly multi-language) , multi-language manual on CD, Waveform Manager Plus software, compact Flash memory card, compact Flash card reader/ writer ( USB connection to PC)

    Options 19-inch rack mounting kit