Specification
8129-01 (for 1ch)
8129-02 (for 2ch)
8129-03 (for 3ch)
� Flexible for various test fields
� Can measure AC current up to 3000A
� Flexible and light weight by the air core coil used at the sensor part
� Measurable conductor size:
max ¦150mm
� Designed to meet the international safety standard
IEC 61010-1 CAT.III 600V
Applicable model 5010/5020 (*),
6300, 6310
* When you use the sensors with KEW5010/5020, please check the "Applicable model list" to check if your KEW5010/5020 can be used with the sensors.
8129-01 (for 1ch) 8129-02 (for 2ch) 8129-03 (for 3ch)
Conductor size : max. ¦150mm
Rated current : 300/1000/3000A
Output voltage : 300A Range : AC500mV/AC300A (1.67mV/A)
1000A Range : AC500mV/AC1000A (0.5mV/A)
3000A Range : AC500mV/AC3000A (0.167mV/A)
Accuracy : ±1.0%rdg (45~65Hz)
Phase Shift : within ±1°
Withstand voltage : AC5350V for 5 seconds
300A Range : 30~300A (45~65Hz)
1000A Range : 100~1000A (45~65Hz)
3000A Range : 300~3000A (45~65Hz)
Cable length : Sensor part : approx. 2m
Output cable : approx. 1m
Output connector : MINI DIN 6PIN
Operating temperature & humidity ranges : 0~50ºC, relative humidity 85% or less (no condensation)
Output impedance : 100© or less
Applicable standards : IEC 61010-1, IEC 61010-2-032 CAT.III 600V Pollution degree2,
IEC 61326
Dimensions : 111(L) x 61(W) x 43(D) mm (except for protrusions)
Weight Approx. 410g (8129-01 (for 1ch)) Approx. 680g (8129-02 (for 2ch)) Approx. 950g (8129-03 (for 3ch))
Accessories : Instruction Manual
7199 (Output Cable) x1
9137 (Carrying Case) (8129-01 (for 1ch)) Instruction Manual
7199 (Output Cable) x2
9137 (Carrying Case) (8129-02 (for 2ch)) Instruction Manual
7199 (Output Cable) x3
9137 (Carrying Case) (8129-03 (for 3ch))