KYORITSU 4120A[Mar. 19, 2009 0:13:30]
Specification
• Custom microprocessor controlled for highest accuracy and reliability.
• 3 LEDs for checking correct wiring status.
• 15mA LOOP measurement:
LOOP impedance 2000© range measurement is carried out with low test current (15mA). The current will not cause tripping out involved RCD even the one with the lowest nominal differential current (30mA).
• Model 4120A has the automatic RCD (ELCB) lock device. There is no need to by-pass most passive RCD's at distribution boards when conducting LOOP or PSC tests.
• Direct reading of Prospective Short Circuit Current (PSC) and earth fault current.
• Safe voltage measurement function.
• Large custom digital display readout on all models.
• Measures low loop resistances (resolution of 0.01©)
• Automatic lock-out if test resister overheats.
• Large custom digital display readout.
• Visual indication of reversed phase and neutral wiring at socket.
• Designed to IP54 Rating
• Complies with IEC 61557
Loop Impedance Ranges:
20/200/2000© (20/200© D-LOK)
Loop Impedance
Accuracy : ±2%rdg±4dgt
AC Test Current : 20© 25A 200© 2.3A
2000© 15mA
AC Test Period : 20© (20ms) 200© (40ms)
2000© (280ms)
PSC Ranges : 200A (2.3A 40ms) 2000A (25A 20ms) D-LOK
20kA (25A 20ms) D-LOK
PSC Ranges Accuracy : Consider accuracy of Loop impedance
Voltage : 110V-260V ±2%rdg±4dgt
Operating Voltage : 230V+10% -15% (195V~253V) 50Hz
Safety Standard : IEC 61010-1 CAT.III 300V
IEC 61557-1,-3
IEC 61010-2-031 Pollution Degree 2
IP Rating : IEC 60529-IP54
Dimensions : 185(L) × 167(W) × 89(D)mm
Weight : 960g approx.
Accessories : Molded plug test leads
7121 (Distribution board test leads)
Pouch for test leads
Shoulder Strap
Instruction Manual