JENCO pH, ORP, Temperature In-line Controller Model : 6303C[Dec. 16, 2010 9:25:54]
Description
1/ 4 DIN panel mount pH controller, microprocessor based with membrane type keypad and password protection. Meter has a programmable, reversible and isolated 4- 20 mA current output and dual SPDT relays. Connectors are a BNC and spade lug terminal blocks. ATC is a 10K thermistor or simulation resistor. Features a NEMA-4 type bezel for watertight seals against mounting panels. 115/ 230V AC.
⢠High accuracy, reversible & isolated 4-20 mA current out
⢠Set point security code protection
⢠Dual output relays
⢠Automatic/ manual temp. compensation
⢠Dual output relays
⢠Automatic/ manual temp. compensation
⢠User programmable buffer recognition or pre-stored nbs buffers
⢠Low drift & high stability
⢠User friendly keypad format
⢠CE approved version available
⢠Power down memory backup
⢠1/ 4 DIN housing
⢠User programmable relay hysteresis
⢠Nema 4 bezel
Range :
pH : -2.00 to 16.00 pH / 0.01 pH / ± 0.01 pH
ORP : -1999 to 1999mV / 1 mV / ± 0.1%
Temp. : -3.0 to 115.0 ° C / 0.1 ° C / ± 0.2 ° C
General
⢠Input impedance: > 10exp( 13) ohms
⢠Temp. compensation: Auto/ manual 0 to 115.0 ° C
⢠Electrode buffer recognition: 4.00, 6.86, & 10.01
⢠Single point standardization: Eliminates electrode offset
⢠Dual point standardization: Calculates/ compensates for electrode solpe & offse
⢠Bumpless calibration: Controller output does not change during calibrati
⢠Display: 0.5 inch high LCD
⢠Lock: Set values connot be changed duless lock code is e
⢠Relay set point hysteresis: User programmable
⢠Relay output ( resistive load) : 5amp. at 115 VAC, 2.5 amp. at 230 VAC
⢠CE certification: Effective january, 1998
⢠NEMA 4: Front of the mounting panel
⢠Power failure backup: Yes
⢠Dimensions: 1/ 4 DIN case
⢠Weight: 680 g
⢠Power source: 115 VAC, 230 VAC ± 15% 50/ 60 hz