Specification
Specifications
Compatible Syringes
10, 20, 30 and 50 mL ( TERUMO and specified brands)
Range of Flow Rate Setting
0.1-300 mL/ h ( for 10, 20 and 30 mL syringes)
0.1-999 mL/ h ( for 50 mL syringes)
0.1-1200 mL/ h ( special function for 50 mL syringes)
Flow Rate Display Change
0.1 mL/ h steps ( for settings from 0.1 to 99.9 mL/ h)
1 mL/ h steps ( for settings from 100 to 999 mL/ h)
Flow Rate Accuracy
Mechanical Accuracy: ± 1%
Accuracy including syringe: ± 3%
Occlusion Detection Pressure
Selectable from three levels.
( Values are with use of TERUMO syringe)
[ III] : 800 ± 200 mmHg max. ( 106.7 kPa ± 26.7 kPa)
[ II] : 500 ± 100 mmHg max. ( 66.7 kPa ± 13.3 kPa)
[ I] : 300 ± 100 mmHg max. ( 40 kPa ± 13.3 kPa)
Optional: 5 occlusion levels ( 100, 200, 300, 500 and 800 mmHg)
Priming ( purge rate)
300 mL/ h ~ 1200 mL/ h ( Depending on syringe size)
Alarms
Occlusion, Nearly empty, Low Battery, Plunger/ Clutch disengaged,
Syringe disengaged, AC/ DC Cable disconnected.
High flow warning when turning the dial over 10 mL/ h ( optional)
Special Functions
Functions are selectable via internal switches:
• Delivery Limit setting
0.1 – 999.9 mL ( in 0.1 mL steps)
When the limit is reached, an alarm will occur and the infusion will
continue at KVO rate of 0.1 mL/ h.
• Nearly Empty alarm setting
The time until the syringe is empty can be programmed.
( from 3 to 60 min.)
• Max. Flow Rate
1 – 1200 mL/ h ( in 1 mL steps)
• Syringe Brand Confirmation
• Memory function ( last settings)
• Changeable flow rate during infusion
Power Supply
100 to 240 VAC, 50/ 60 Hz
12 to 15 VDC
Operable duration using internal battery NiMH:
Approx. 5 hours ( using a new battery at flow rate of 5 mL/ h and
ambient temperature of 25° C, after charging at least 15 hours with
the power off)
Classification ( acc. to IEC 60601-1)
Class I and internally powered equipment, Type CF, Continuous
operation, IPX4 ( splashproof)
External Dimensions
322 ( W) x 114 ( H) x 115 ( D) mm
Weight
TE-331 ± 1.8 kg
TE-332 ± 1.9 kg
External Communication and Nurse Call Connector