Flue Gas CO, O2 Analyzer Model : GCO-200 Brand DKK-TOA
Specification
Flue Gas CO, O2 Analyzer Model GCO-200
Partial revision of the Ordinance for nforcement of the Waste Management and Public Cleansing Act ( put in e f f e c t i n D e c e m b e r 1 9 9 7 ) , i t i s n o w m a n d a t o r y t o continuously measure and record CO concentration in the exhaust of refuse incinerators and, when incinerating
refuse, limit CO concentration in the exhaust to no more than 100 ppm.
FEATURES :
° Easy to maintain, install and move because of simple structure.
T h e m a i n m a i n t e n a n c e w o r k i n c l u d e s r e g u l a r
replacement of filters ( once every two weeks to one month) and standard gas calibration ( once in three m o n t h s ) . I t u s e s a c o m p a c t s e n s o r w h i c h , w h e n broken, can be easily replaced.
° Excellent in long-term stability
In ordinary use, the system requires no standard gas calibration for over three months.