Specification :
Features:
--Preheating counter: 4 pieces ( for double-channel) , 1 piece ( for single channel) , prompt sound arriving time.
--Permanently store fixed standard curve.
--linkage adding sampler, automatic touching-off testing.
--LCD crystal, Chinese display, man-machine dialogue operation, easy to learn.
--Built-in temperature sensing printer, printing single sample, single item and universe report.
--Full open program, arbitrary choice for reagent.
--Relative movement of steel ball is not be influenced by original blasma viscosity.
--The tested plasma is not interfered by icterus, hemolysis, chle, turbidity, etc.
Main technical:
--Testing item( 18 items) : PT, APTT, FIB, TT, HEP, LMWH, albumen C, albumen S, SPA, ATIII, blood coagulation factor II, V, X, VIII, IX, XI, XII
--Measuring accuracy: ± 3%
--Repeatability: ± 3%
--Linearity precision: R> 0.99
--Temperature control precision: 37° C± 0.3° C