OCMA-350 OIL CONTENT ANALYZER MERK HORIBA Hub : 021 9600 4947, 0815 7477 4384[Feb. 5, 2012 10:28:37]
Overview
Just by setting the equipment, the OCMA-350 provides highly accurate analysis of the oil content of water samples from which the oil has already been extracted in the auxiliary cell. In addition to displaying residual oil and solid body content suspended in water in a range from 0mg/ l-200mg/ l, the OCMA-350 allows users to display analysis results for land in mg/ kg and for oil absorbency in Abs. The OCMA-350 is also well suited for evaluating the capabilities of semiconductor fabrication and precision machinery cleaning equipment, evaluating the properties of industrial process oil and the residual oil of polishing materials as well as wastewater that has been adulterated with silt, sludge, and other suspended particles.
OCMA-350 literature
Features
* The unit provides swift analysis using a solvent extraction/ infrared absorptiometry method.
* The units uses a desorption cell and is designed for easy sample injection.
* Oil is extracted using the extremely stable SS-316 solvent extractant.
* Users can display analysis results in three different units-mg/ l, mg/ kg, and Abs.
* The unit features an easy-to-read backlit liquid crystal display.
* The units feature an RS-232C interface as standard equipment.
Measurement principle:
Solvent extraction, NDIR
analysis ( Infrared spectroscopy)
Specifications :
Measurement range:
0 to 200 mg/ L, 0 to 1000 mg/ kg, 0 to 1 Abs.
Resolution
mg/ L;
0 to 99.9 mg/ L; 0.1 mg/ L
100 to 200 mg/ L; 1 mg/ L
mg/ kg;
0 to 9.99 mg/ kg; 0.01 mg/ kg
10.0 to 99.9 mg/ kg; 0.1 mg/ kg
100 to 1000 mg/ kg; 1 mg/ kg
Abs;
0 to 1000 Abs; 0.001 Abs
Repeatability
mg/ L;
0 to 9.9 mg/ L; ± 0.4 mg/ L ± 1 digit
10.0 to 99.9 mg/ L; ± 2.0 mg/ L ± 1 digit
100 to 200 mg/ L; ± 4 mg/ L ± 1 digit
Abs;
± 1% of full scale
Measurement
Manually controlled
Calibration
Automatic calibration after the calibration standard is introduced to the instrument
Extraction Solvent
S-316 solvent
Sample/ Solvent Volume
Approx. 6.5 mL, ratio of sample to solvent is 1: 1
Extraction Method
Built-in extractor for oily water samples
Display
Measured value; 3 1/ 2digits LCD with back-light
Message; Character display LCD with back-light ( 16 x 2 characters)
Functions
Self diagnostics, Auto hold function, Calendar clock
Output
RS-232C and centronics printer port
Ambient Temperature
0 to 40
Power
100 to 240 V AC ± 10% , 50/ 60 Hz, 60 VA
Dimensions
200 ( H) x 250 ( W) x 285 ( D) mm
Mass
Approx. 5 kg