  • Products Catalog

    CONDUCTIVITY METERS - THERMOMETERS, Model : HD2106.1 AND HD2106.2, Brand : DeltaOhm

    [Nov. 29, 2010 3:29:40]
    CONDUCTIVITY METERS - THERMOMETERS,  Model : HD2106.1 AND HD2106.2,  Brand : DeltaOhm
    Quantity Order
    Add to Basket
    Origin
    Italy
    Recommend to your friends Put on Product Partner

    Specification

    The HD2106.1 and HD2106.2 are portable instruments with a large LCD display. They measure conductivity, liquid resistivity, total dissolved solids ( TDS) , and salinity using combined 4-ring and 2-ring conductivity/ temperature probes. Temperature only is measured by Pt100 or Pt1000 immersion, penetration or contact probes.
    The probe calibration can be performed automatically in one or more than one of the 147¼ S, 1413¼ S, 12880¼ S or 111800¼ S/ cm conductivity calibration solutions. The temperature probes are fitted with an automatic detection module, with the factory calibration settings already being memorized inside. The HD2106.2 is a datalogger. It memorizes up to 36, 000 conductivity and temperature samples which can be transferred from the instrument connected to a PC via the multi-standard RS232C serial port and USB 2.0. The storing interval, printing, and baud rate can be configured using the menu.
    The HD2106.1 and HD2106.2 models are fi tted with an RS232C serial port and can transfer the acquired measurements to a PC or to a portable printer in real time.
    The Max, Min and Avg function calculates the maximum, minimum or average values. Other functions include: the relative measurement REL, the Auto-HOLD function, and the automatic turning off which can also be disabled.
    The instruments have IP67 protection degree.

    INSTRUMENT TECHNICAL CHARACTERISTICS
    Instrument
    Dimensions ( Length x Width x Height) 185x90x40mm
    Weight 470g ( complete with batteries)
    Materials ABS, rubber
    Display 2x41D2 digits plus symbols
    Visible area: 52x42mm
    Operating conditions
    Working temperature -5… 50° C
    Storing temperature -25… 65° C
    Working relative humidity 0… 90% RH without condensation
    Protection degree IP67
    Power
    Batteries 4 1.5V type AA batteries
    Autonomy 200 hours with 1800mAh alkaline batteries
    Power absorbed with instrument off 20¼ A
    Mains Output mains adapter 9Vdc / 250mA
    Security of memorized data Unlimited, independent of battery charge
    conditions
    Time
    Date and time Schedule in real time
    Accuracy 1min/ month max error
    Measured values storage - model HD2106.2
    Type 2000 pages containing 18 samples each
    Quantity 36000 pairs of measurements [ Ç -° C] ,
    [ © -° C] , [ TDS-° C] or [ Sal-° C]
    Storage interval 1s… 3600s ( 1hour)
    Serial interface RS232C
    Type RS232C electrically isolated
    Baud rate Can be set from 1200 to 38400 baud
    Data bit 8
    Parity None
    Stop bit 1
    Flow Control Xon/ Xoff
    Serial cable length Max 15m
    Immediate print interval 1s… 3600s ( 1hour)
    USB interface - model HD2106.2
    Type 1.1 - 2.0 electrically isolated
    Connections
    Conductivity input 8-pole male DIN45326 connector
    Input module for the temperature probes 8-pole male DIN45326 connector
    Serial interface and USB 8-pole MiniDin connector
    Mains adapter 2-pole connector ( positive at centre)
    Measurement of conductivity by Instrument
    Measurement range ( K cell= 0.1)
    Resolution 0.00… 19.99¼ S7cm / 0.01¼ S/ cm
    ( with K cell= 0.1)
    0.0… 199.9¼ S / cm / 0.1¼ S/ cm
    200… 1999¼ S / cm / 1¼ S/ cm
    2.00… 19.99mS / cm / 0.01mS/ cm
    20.0… 199.9mS / cm / 0.1mS/ cm
    Accuracy ( conductivity) ± 0.5% ± 1digit
    Measurement of resistivity by Instrument
    Measurement range / Resolution 4.0… 199.9© / 0.1©
    200… 999© / 1©
    1.00k… 19.99k© / 0.01k©
    20.0k… 99.9k© / 0.1k©
    100k… 999k© / 1k©
    1… 10M© / 1M©
    Accuracy ( resistivity) ± 0.5% ± 1digit
    Measurement of total dissolved solids ( with coefficient Ç / TDS= 0.5)
    Measurement range ( K cell= 1) 0.00… 19.99mg/ l / 0.05mg/ l
    Resolution 0.0… 199.9 mg/ l / 0.5 mg/ l ( with K cell= 0.1)
    200… 1999 mg/ l / 1 mg/ l
    2.00… 19.99 g/ l / 0.01 g/ l
    20.0… 199.9 g/ l / 0.1 g/ l
    Accuracy ( total dissolved solids) ± 0.5% ± 1digit
    Measurement of salinity
    Measurement range / Resolution 0.000… 1.999g/ l / 1mg/ l
    2.00… 19.99g/ l / 10mg/ l
    Accuracy ( total dissolved solids) ± 0.5% ± 1digit
    Measurement of temperature by Instrument
    Pt100 measurement range -50… + 200° C
    Pt1000 measurement range -50… + 200° C
    Resolution 0.1° C
    Accuracy ± 0.25° C
    Drift after 1 year 0.1° C/ year
    Automatic/ manual temperature
    compensation 0… 100° C with ± T= 0.00… 4.00% / ° C
    Reference temperature 20° C or 25° C
    Ç / TDS conversion factor 0.4… 0.8
    Cell constant K ( cm-1) 0.1, 0.7, 1.0 and 10.0
    Standard solutions
    automatically detected ( @ 25° C)
    147 ¼ S/ cm
    1413 ¼ S/ cm
    12880 ¼ S/ cm
    111800 ¼ S/ cm