Welcome to Indomonster.com
Sign In | Sign Up

Welcome to

MIKASE LABORATORY - PT MITRA KATIGA SEJAHTERA / SUPPLIER ALAT UJI KUALITAS UDARA, AIR, CUACA , KEBISINGAN SERTA GETARAN

http://indomonster.com/mikaselaboratory/

Priority Member Indomonster Priority
16th Year

showing 116 of 191
  • Products Catalog

    Transmitter, indicator , ON/ OFF regulator , temperature and humidity datalogger with interchangeable probe, Model : HD 2717T, Brand : DeltaOhm

    Quantity Order
    Add to Basket
    Origin
    Italy
    Recommend to your friends Put on Product Partner

    Specification

    The instruments of the HD2717T... series are transmitters, indicators, and ON/ OFF regulators with data logging functions, they measure temperature and humidity.
    The main feature of these instruments is their interchangeable probe. The probe can be replaced by the user without process interruption. Thus, the probe can be calibrated or repaired at a later time.
    The instrument is available in three different versions: with horizontal probe ( S.TO) , vertical probe ( S.TV) or with remote probe ( S.TC) , having the probe connected to the electronics by means of a cable of various lengths. The S.TO and S.TV probes are made of stainless steel AISI304, the S.TC probes can be of stainless steel AISI304 or POCAN ( plastic material) .
    The probe is factory calibrated and ready to use, it is provided with a SICRAM2 module which
    stores the calibration data of the probe, allowing the interchangeability of the probes.
    The instruments measure:
    • Temperature in Celsius or Fahrenheit temperature scale
    • Relative humidity
    and calculate:
    • Absolute humidity
    • Mixing Ratio
    • Dew point
    All models have both current and voltage outputs.
    Some models are fitted with two control relays and one alarm relay, configurable by the user.
    All models are fitted with a multistandard RS232/ RS485 serial port and an auxiliary RS232C standard serial output. The RS485 serial output allows the management of more than one device in a network.
    The models HD2717T... can be with or without LCD. The display shows on the first line the relative humidity or a derived parameter and on the second line the temperature in degrees Celsius or Fahrenheit.
    The data logger function allows to store the measures with a selectable storage interval.
    The instrument setup remains permanently stored, while the real time clock is protected by an apposite Lithium battery against temporary mains voltage interruptions.
    The power supply can be chosen, at the time of placing the order, between 24Vac/ dc or
    universal 90...240Vac.

    Technical specifics ( @ 24Vac and 20° C)
    Inputs
    Temperature Sensor Pt100 classe 1/ 3 DIN
    Working range of the sensor
    -50 ... + 200° C ( -58 ...+ 392° F)
    Humidity Relative humidity % RH 5 ... 98% RH
    Working range of the sensor in temperature
    -50 ... + 150° C
    ( Special configurations up to 180° C available on request)
    Dew point TD -50 ... + 100° C
    Absolute humidity 0 ... 600g/ m3
    Mixing ratio 0 ... 2000g/ kg of dry air
    Wet bulb temperature -50 ... + 100° C
    Accuracy of the measured physical quantities
    Temperature Pt100 ± 0.25° C
    Relative humidity % RH
    ± 2.5% RH ( 10...90% RH)
    ± 2.5% RH in the remaining range
    Accuracy of the calculated
    physical quantities
    See table in the following chapter
    Response time
    3min with grid protection ( at 20° C and 0.5m/ s)
    Outputs
    Communications Type RS232C and RS485 Multidrop
    Baud Rate
    9600 baud
    57600 baud non-permanent
    Physical quantities Measured Temperature, relative humidity
    Calculated
    Dew point, absolute humidity,
    ( mixing ratio) .
    Analog
    outputs
    Output types
    4...20mA; 0...20mA
    0...10Vdc; 2… 10Vdc
    Load resistance
    Current output: 500© max
    Voltage output: 100k© min
    Resolution 16bit
    Accuracy analog outputs ± 0.05% f.s. @ 20° C
    In case of measuring error
    ( exceeding of the operating
    limits, faulty or not connected
    probe, ....)
    Idc = 22mA
    Vdc = 11V
    Relay Working relay
    2 x 3A/ 250Vac Load resistance,
    1 change-over contact
    Alarm relay
    1 x 3A/ 250Vac Load resistance,
    1 with normally open contact
    Instrument
    Power supply Versions
    24Vdc / 24Vac 50… 60Hz, ± 10%
    90 ... 240Vac 50… 60Hz
    Average consumption 3W
    Data logger Storage capacity 9000 samples in max. 256 sessions
    Storage type Circular memory
    Stored parameters
    Temperature, relative humidity, dew point, absolute humidity, mixing ratio, analog outputs 1 and 2, relay status1, 2, 3.
    Storage interval
    1, 2, 5, 10, 20, 60 seconds, 2 and 4 minutes
    Real time clock Type
    Real time with Lithium buffer battery
    Accuracy ± 1min/ month
    Software DeltaLog12
    for Windows® 98 to Vista operating systems
    Display LCD Custom segment LCD
    Ambient working conditions of the electronics
    Operating temperature -20… + 60° C
    Relative humidity 0… 90% RH - No condensate
    Static working pressure of the sensors 12 bar max.
    Storage temperature -30… + 80° C
    Housing L x H x W 143x154x61
    Weight 600g
    Material ABS
    Degree of protection Electronics IP65