Specification :
Applications:
This instrument is used for analyzing & measuring the content & concentration of elements ( K, Na, Li, Ca, Ba) . It can be widely applied to laboratory, clinical experiments and pathology research ( serum, urine, plasma) , agriculture ( soil, fertilizer, tobacoo) , industry ( cement, glass, ceramics, mineral or ore powder, fireproof material) , food ( wine, water) , etc.
Features:
# K & Na filters & air-compressor as standard accessories.
# Li, Ca, Ba filters as optional accessories.
# Dual channel & 4 digital LED display.
# Unique ignition path & keyboard operate.
# Electronic flame failure detection.
# Fine & coarse sensitivity controls.
# Operates on L.P.G supply.
# Easily maintained.
Specifications
Model: FP6400
Measurement Range: K: 0.0-1.0mmol/ L, Na: 0.0-5.0mmol/ L.
Detection Limit: Kd0.004mmol/ L, Nad0.008mmol/ L.
Linearity Range: K: 0.02mmol/ L-0.1mmol/ L, Na: 0.1mmol/ L-0.5mmol/ L.
Linearity Error: Kd0.005mmol/ L, Nad0.03mmol/ L.
Repeatability: Relative Standard Deviationd3% .
Stability: Within 15s the indication of the Max. Relative Deviation d3% .
Dimension: 410× 675× 410mm.
Weight: 10 kg
Model: FP640
Measurement Range: K: 0.0-2.0mmol/ L, Na: 0.0-10.0mmol/ L.
Detection Limit: Kd0.004mmol/ L, Nad0.008mmol/ L.
Linearity Range: K: 0.02mmol/ L-0.1mmol/ L, Na: 0.1mmol/ L-0.5mmol/ L.
Linearity Error: Kd0.005mmol/ L, Nad0.03mmol/ L.
Repeatability: Relative Standard Deviationd3% .
Stability: Within 15s the indication of the Max. Relative Deviation d3% .
Dimension: 410× 675× 410mm.
Weight: 10 kg
Model: FP6410
Measurement Range: K: 0.0-5.0mmol/ L, Na: 0.0-25.0mmol/ L.
Detection Limit: Kd0.004mmol/ L, Nad0.008mmol/ L.
Linearity Range: K: 0.02mmol/ L-0.1mmol/ L, Na: 0.1mmol/ L-0.5mmol/ L.
Linearity Error: Kd0.005mmol/ L, Nad0.03mmol/ L.
Repeatability: Relative Standard Deviationd3% .
Stability: Within 15s the indication of the Max. Relative Deviation d3% .
Dimension: 410× 675× 410mm.
Weight: 10 kg
Model: FP6420
Measurement Range: K: 0.0-10.0mmol/ L, Na: 0.0-200.0mmol/ L.
Detection Limit: Kd0.004mmol/ L, Nad0.008mmol/ L.
Linearity Range: K: 0.02mmol/ L-0.1mmol/ L, Na: 0.1mmol/ L-0.5mmol/ L.
Linearity Error: Kd0.005mmol/ L, Nad0.03mmol/ L.
Repeatability: Relative Standard Deviationd3% .
Stability: Within 15s the indication of the Max. Relative Deviation d3% .
Dimension: 410× 675× 410mm.
Weight: 10 kg