Specification :
Horizontal-Vertical Flame Chamber Tester ZLT-UL94
Product details: Model ZLT-UL94: The ZLT Horizontal-Vertical Flame Chamber Tester is a fully automatic instrument contained in its own cabinet to maximize the safety of the operator. It screening procedure for comparing the relative burning behaviour specimens made from plastic and other non-metallic materials , exposed to a flame ignition source of 50W or 500W nominal power. These test methods determine the linear burning rate and the afterflame/ afterglow times, as well as the damaged length of specimens.They are applicable to solid and cellular materials that have an apparent density of not less than 250kg/ m3, determined in accordance with ISO845.Then do not apply to materials that shrink away from the applied flame without ignitiong; ISO9773 should be used for thin flexible materials. And it' s the combustion test device of plastic materials represented by UL-94. It can perform each test of HB, V0 to V2, 5V and VTM, HBF.
Product three type: 50W 500W 50~ 500W mixed-type test equipment
SPECIFICATIONS
Test Flame: 50W( burning test for V-0, V-1, V-2 classification)
Burner Tube: Internal diameter: D 9.5mm± 0.3mm, length: 100mm± 10mm
Burning Angle: 0° , 20° , 45°
Thermocouple: Type K ( Ni/ Cr - Ni/ Al)
Thermocouple Size O/ D: 0.5 mm
Standard Copper Block: $ 5.5mm± 0.01mm, weight: 1.76g± 0.01g before drilling. Cu-ETP
Time for Temperature to Increase: From 100° C± 5° C to 700° C± 3° C; 44s± 2s ( preset)
Gas Supply: 18mm~ 22mm( adjustable)
Flowmeter: methane, min.98% purity
Back Pressure: 105ml/ min± 2%
Flame Height Indicator: Less than 10mm of water
Input Power: 240 Vac 50 Hz 3A or 115 Vac 60Hz Optional
Volume Of Chamber: > 0.5m3, black interior
Outer Dimensions: W* D* H= 1100mm* 550mm* 1100mm
Conform to standard: UL94, IEC60695-11-4, IEC60695-11-10
Fitted to the base plate are a flame height scale, a level and adjustable feet.
DIGITAL TEMPERATURE DISPLAY
Includes: 50W Flame height gauge and 100% absorbent cotton