Specification :
Characteristics
Throughput: 200 test /hour
Test mode:
End –point, Kinetics, Fixed time
Single and Double wavelength
Linearity and Non-linearity calibration
With and without reagent blank and (or) sample blank
Can use Single reagent and double reagent
Test range: clinical biochemistry, turbidimetry and immunoassay tests
Setting of test: single parameter test, multi-parameters test, combined parameters, batch parameters test
Setting of parameters:
Completely open for editing test parameters, editing combined parameters and calculating parameters
Bench top compact automatic analyzer
Routine & STAT tests available
User-friendly, safe handling
Technical Parameters
Optical system
Light source: 12V/20W Halogen lamp, life span>2000h
Wavelength: 11 interference filters: 340, 380, 405, 450, 492, 505, 546, 578, 620, 670 And 700nm
Sample system
Sample plate: two rotor rings,64 positions for samples, all positions can be assigned as sample, standard, quality control and blank
Sample volume: 5^50ul
Sample auto-dilution: instrument dilutes and retests the sample automatically
Sample needle: Auto liquid level detection function; auto-wash inner and outer wall
Reagent system
Sample plate: 30 reagent positions, all positions can be assigned as R1, R2
Reagent bottle volume: 30ml
Reagent refrigeration and heating: 4^10 Crefrigeration and reagent preheating functions
Reagent volume: 30ml
Reagent needle: Auto liquid level detection function; auto-wash inner and outer wall
Reaction system
Reaction plate: turnable reaction plate,96 disposable reaction cuvettes
Reaction Temperature: 37, temperature accuracy is ±0.1
Quality control:
3 control sera levels, calculating statistical results automatically, levey-jennings QC plot. Real-time inquiry the QC, print QC results and QC plot
Data system:
Computer control: Windows XP system
Data processing: large capacity can store test results, display, save and print reaction curve
Print: edit and print the completed English report, can choose and edit the printing content, support the net work report printing
Working environment:
Power supply: AC220V±10%,50Hz±2%
Electric power consumption: