Specification :
DIGITAL CLAMP ON HiTESTER 3282
The true RMS is shown in the distorted waveform
AC Current/ Voltage, Frequencies, Resistance, Conduction check, Temperature ( probe sold separately)
Wave peak value, Waveform distortion check
Auto-power off to prevent the power from being left on
Non-fuse type protects up to 600 V AC
Basic specifications
AC Current range 30.00 to 1000 A AC, 3 ranges ( ± 1 % rdg. ± 5 dgt. at 50 or 60Hz) , True RMS
Wave peak value at AC Current 30.0 to 600 A peak, 3 ranges ( ± 3 % rdg. ± 5 dgt. at 50 or 60Hz)
AC Voltage range 300.0 or 600 V AC, 2 ranges ( ± 1 % rdg. ± 3 dgt. at 50 or 60Hz) , True RMS
Wave peak value at AC Voltage up to 300 or 600 V peak, 2 ranges ( ± 3 % rdg. ± 5 dgt. at 50 or 60Hz)
Other functions
Distortion check: 1 to 5 Crest factor, Resistance: 1 k or 10 k-ohm range, Temperature: -50 to 150 ° C ( TEMPERATURE PROBE 9462 required, sold separately) , Frequency: 100 or 1000 Hz, Mode: Slow/ Peak/ C.F./ RMS/ Record mode/ Auto-off/ Conduction
Analog output None
Frequency characteristics
( at AC current / voltage) 40 to 1000 Hz
Display section Digital / 3000 dgt. Bar graph / 35 seg.
Sampling rate 2 or 4 times/ sec ( Slow: 1 times/ 3 sec)
Crest factor ( RMS) 2.5 ( 1.7 at 1000 A, 600 V range)
Max. circuit voltage 600 V AC rms ( insulated wire)
Power supply 6F22( 006P) × 1, Continuous use 45 hours
Core jaw dia. Ø 46 mm( 1.81 in)
Dimensions, mass 62 mm( 2.44 in) W × 230 mm( 9.06 in) H × 39 mm( 1.54 in) D, 400 g ( 14.1 oz)
Accessories TEST LEAD 9207-10 ( 1) , CARRYING CASE 9399 ( 1) , Hand strap ( 1)